Search Results for author: O. Sato

Found 2 papers, 0 papers with code

Development of New Tracking Detector with Fine-grained Nuclear Emulsion for sub-MeV Neutron Measurement

no code implementations29 Jan 2021 T. Shiraishi, I. Todoroki, T. Naka, A. Umemoto, R. Kobayashi, O. Sato

First, we calibrated the performance of NIT device for detecting monochromatic neutrons with sub-MeV energy generated by nuclear fusion reactions, and the detection efficiency for recoil proton tracks of more than 2 um range was consistently 100\% (the 1 sigma lower limit was 83%) in accordance with expectations by manual based analysis.

Instrumentation and Detectors